SpinCore Logo Home
Applications
Contact Us
Purchasing Info
About Us
Software Downloads

banner_left.jpg banner_right.jpg




T1 Inversion-Recovery Program

for RadioProcessor and iSpin-NMR


Overview

The T1 Inversion-Recovery (T1IR) experiment applies a 180 degree inverting pulse, waits a period of tau, then applies a 90 degree pulse to cause an FID response from the sample. It then repeats this pulse sequence with increasing durations of tau. The relationships of the FID response peaks can be modeled with an exponential fit to determine T1, the Spin-Lattice relaxation time.

The length of the T1IR experiment can make it easy for the magnet and sample to drift off resonance during experimentation and skew results. To prevent this, the T1IR program has the ability to use a Single-Pulse NMR scan to automatically find and set the resonance frequency before each iterative Inversion Recovery scan. This auto-resonance feature will increase run time, but prevent any off-resonance drifting.

The T1IR program is available for download at the bottom of this page.


Figure1. T1IR Manual Range Results (16 tau steps, tau range: 0.1ms - 300ms, a sample of household cooking oil using a 10.8MHz permanent magnet)


Figure 1 shows the resultant FID peak response when manually ranging tau. Figure 2, below, shows the pulse sequence used by the RadioProcessor to perform this experiment. Each vertical line is a new instruction, and the names of the durations used are shown above each time range.



Figure2. T1IR Pulse Sequence



Running the T1IR Program

The T1IR program accepts command line parameters to specify the values for each of the pulse program's parameters. It takes the following arguments as inputs:


Argument
Parameter
Units/Values
1
File Name

2
Board Number
#
3
De-blank Bit
0, 1, 2, or 3
4
Debug 0 or 1
5
ADC Frequency
MHz
6
Spectrometer Frequency
MHz
7
Spectral Width
kHz
8
Amplitude
0.0 - 1.0
9
90 Degree Pulse Time
us
10
90 Degree Pulse Phase
degrees
11
180 Degree Pulse Time
us
12
Bypass FIR
0 or 1
13
Number of Points
#
14
Number of Averaging Points
#
15
Number of Scans
#
16
Tau Steps
#
17
Auto Tau Range
0 or 1
18
Auto Tau Range Seed
ms
19
Step Distribution
#
20
Minimum Tau
ms
21
Maximum Tau
ms
22
De-blanking Delay
ms
23
Transient Delay
us
24
Repetition Delay
s
25
Correct DC Offset
0 or 1
26
Auto Phasing
0 or 1
27
Auto Resonance
0 or 1
Table1. T1IR Input Arguments


These arguments must all be passed as input parameters when launching the T1IR.exe file. Omitting or passing arguments out of order will cause the program to return an error. To simplify this process, a T1IR.bat file is provided. This file allows for easy manipulation of argument values, and ensures that they will passed in the correct order. To use the T1IR.bat file simply:
  • Open the T1IR.bat file in a text editor
  • Set the parameters to the desired values
  • Save any changes made
  • Ensure that the T1IR.bat and T1IR.exe files are in the same directory
  • Double-click the T1IR.bat file
After following these procedures, the T1IR.exe will run with the specified parameters and print those parameters as well as the progress of the experiment to a terminal window. Once experimentation is finished, it will save the results to ASCII, JCAMP, Felix, and gnuplot files. If gnuplot has been installed, the T1IR.bat file will automatically run gnuplot with the outputted gnuplot file. This will calculate a value for T1 as well as display a plot of the data.



Viewing T1IR Results

After execution of the T1IR program, the output files can be used with external programs to visualize the data. The T1IR program wills

The LabVIEW NMR Interface can process .txt files. To do this:
  • Navigate to the 'Process Data' tab
  • Point the 'Import File Name' path to the T1IR's .txt output.
  • Press 'Import Data From ASCII File' to view the data.


The Felix NMR program can process .fid files. To do this simply ensure that the Felix NMR program is installed properly and double-click on the .fid file.


Gnuplot can process .gp files. To do this:
  • Download and install the latest version of gnuplot.
    • Use the default installation settings.
    • To be able to automatically open gnuplot with the .bat file, make sure to check 'Add application directory to your PATH environment variable' in the 'Select Additional Tasks' window.
  • If gnuplot has been setup to be called in the terminal, results should automatically be viewed after each execution of the .bat.
  • If for whatever reason gnuplot is not running from the terminal, results can still be viewed by running the .gp file in the dedicated gnuplot GUI. To do so:
    • Identify the directory containing T1IR.exe and T1IR.bat. This can be found in the T1IR.bat's properties window:



    • Launch gnuplot.exe.
    • Point gnuplot to the directory by inputting the following command to the gnuplot terminal:
      cd '<directory address>'



    • Call the gnuplot script with the following command:
      call '<T1IR_filename>/<T1IR_filename>_plot.gp'



    • View the generated T1 value and output .jpg.



T1IR Acquisition Tips

Setting the input parameters correctly is critical in obtaining good T1IR data. Below are suggestions for determining these parameters:
  • Start with a simple Single-Pulse NMR experiment:
    • Find the values of spectral width and number of points which optimize the signal-to-noise ratio without cutting off the FID.
    • Find the 90 and 180 degree pulse widths. These are the amounts of time it takes for a transmitted pulse to cause a 90 and 180 degree FID response. The 180 degree pulse width should be roughly double that of the 90 degree pulse width.
    • Find the resonance frequency and output phase which achieve a maximum real amplitude at the start of acquisition. The value of the first real data point should be a maximum, and the value of the first imaginary point should be zero.
    • These procedures are done using the LabVIEW NMR Interface.
  • Use the established Single-Pulse NMR parameters with the T1IR program. The relevant parameters are:
    • BOARD_NUMBER
    • ADC_FREQUENCY
    • SPECTROMETER_FREQUENCY
    • SPECTRAL_WIDTH
    • AMPLITUDE
    • P90_TIME
    • P90_PHASE
    • P180_TIME
    • TRANSIENT_DELAY
    • REPETITION_DELAY
  • Do an initial run of T1IR. Start with a broad range of TAU values and use a number of TAU_STEPS of around 16 points.
  • Using this initial value of T1, edit the TAU range by setting TAU_MAX to around 4*T1.
  • Finally, adjust the number of TAU_STEPS as desired to get more or less curve resolution. You can also increase the NUMBER_OF_SCANS parameter to average results over multiple runs. This will improve signal-to-noise ratio.
  • If the results appear to be drifting off resonance during iterative scans, make sure to set AUTO_RES to 2. This will ensure that each iterative T1IR scan will be on resonance.


Download



Home | Products | Applications | Contact Us | Purchasing Info | About Us | Software Downloads

© 2022 SpinCore Technologies, Inc.